Medical Isolation Monitoring Solution
1. Standard Compliance & Application Background
- Operating Rooms (OR)
- Intensive Care Units (ICU)
- Neonatal Intensive Care Units (NICU)
- Cardiac Catheterization Labs (DSA/CCU)
- Surgical Recovery Rooms
- Intermediate Care Units

2. Full Product Portfolio of Acrel Medical IT System
2.1 AITR Series Medical Isolation Transformer
- Rated power options: 3.15kVA / 5kVA / 6.3kVA / 8kVA / 10kVA
- Built-in dual series PT100 temperature sensors for overheat monitoring
- H-class reinforced insulation, withstand 4200V/min dielectric voltage test
- Ultra-low leakage current <0.2mA, efficiency over 96%, noise level <40dB
- Low inrush current (<12 times rated current) to avoid surge impact on medical equipment

2.2 AIM-M Series Insulation Monitoring Devices (IMD)
- AIM-M10 (Economic Model)
DIN rail mounted (54×85×64mm), basic monitoring for small operating rooms without fault location.
- Real-time insulation resistance detection (10~999kΩ adjustable threshold)
- Transformer load current & winding temperature monitoring
- Self-test function, PE line breakage alarm, 1 relay output
- RS485 Modbus-RTU communication, stores up to 20 fault records

2. AIM-M300/SG (Advanced Model)
Supports matching with AIL fault locators for circuit pinpointing, suitable for ICU/NICU multi-bed wards.
- Retains all functions of AIM-M10
- Built-in locating current injector for insulation fault positioning
- Dual communication ports: RS485 Modbus + CAN bus
- Two relay alarm outputs for on-site and nurse station notification

2.3 AID Series Alarm & Display Panels

2.4 AIL Series Insulation Fault Locators
- Models: AIL150-4 (4 channels), AIL150-8 (8 channels, max 24 channels per IT system), AIL160-6 (6 channels, max 48 channels per IT system)
- DC 9~36V wide power supply, CAN bus communication
- Compliant with IEC 61557-9 standard, CE & IEC certified
- Independent current transformers for each circuit to detect fault branch rapidly

2.5 Auxiliary Supporting Components
- AKH-0.66P26 Current Transformer: Collect load current signal for IMD measurement, 0.5% high accuracy, 5000V isolation voltage
- HDR-60-24 DC Power Supply: DIN rail installation, 100~240V AC input, 24V DC 2.5A output, supplies power to AIM-M300, AID and AIL devices

3. Three Standard System Schemes for Different Medical Zones
Scheme 1: Basic Operating Room Solution (No Fault Location)
Suitable for single operating room with limited budget
Bill of Materials:
1 × AITR8000S Medical Isolation Transformer
1 × AIM-M10 Insulation Monitoring Device
1 × AKH-0.66P26 Current Transformer
1 × AID10 / AID150 Alarm Display Panel

Scheme 2: Advanced Operating Room Solution (No Fault Location)
Suitable for large operating rooms with stable power supply monitoring demand
Bill of Materials:
1 × AITR8000S Medical Isolation Transformer
1 × AIM-M300 IMD
1 × AKH-0.66P26 Current Transformer
1 × HDR-60-24 DC Power Supply
1 × AID150 Alarm Display

Scheme 3: ICU / NICU Comprehensive Solution (With Fault Location)
Mandatory for multi-bed intensive care wards requiring rapid fault troubleshooting
Bill of Materials:
1 × AITR8000S Isolation Transformer
1 × AIM-M300/SG IMD with locating signal generator
1 × AKH-0.66P26 Current Transformer
1 × HDR-60-24 Power Supply
1 × AIL160-6 / AIL150 Series Fault Locator
1 × AID150 / AID200 Central Display Panel (Installed at Nurse Station)

4. System Topology for Hospital Power Distribution
The whole hospital power framework adopts dual mains supply + backup diesel generator. Power is divided into Group 0, Group 1 and Group 2 medical loads after low-voltage distribution.
All Group 2 critical medical areas (OR, ICU, NICU, DSA room) are equipped with independent Acrel medical IT isolation systems, each set with dedicated UPS backup to guarantee zero power interruption during insulation single fault.

5. Certifications & On-site Project Cases
Global Authoritative Certifications
- IEC 60364-7-710:2021
- IEC 61557-8 / IEC 61557-9
- IEC 61558-1 / IEC 61558-2-15

Overseas Project References
- Saudi Arabia, Romania, Brazil, Russia, Morocco, Malaysia, Pakistan
On-site installation photos include integrated isolation power cabinets, wall-mounted display panels, complete low-voltage distribution boxes inside operating rooms and ICU equipment rooms.

6. Core Advantages of Acrel Medical IT Solution
- Full Standard Compliance: Strictly follows IEC international medical electrical standards, meets local hospital bidding requirements in Europe, Middle East, Southeast Asia and Latin America
- Modular Customization: Three graded schemes for different medical areas, flexible matching according to ward scale and budget
- High Safety Performance: Ultra-low leakage current, high insulation level, dual temperature protection of transformer, double fault alarm mechanism
- Intelligent Centralized Monitoring: Supports RS485/CAN communication, compatible with AcrelEMS-Med hospital energy management system for remote cloud monitoring
- Easy Maintenance: Optional automatic circuit fault localization, reduces troubleshooting time for engineering teams
- Global After-sales Support: Localized service teams in target overseas markets, complete product documentation and technical PPT materials for customers